Lab Apparatus
Variac
A variable AC power supply. Input is connected to
Rheostat
A variable resistor. Used to control the current in a circuit.
Multimeter
Measures voltage, current, frequency.
There are 3 variants:
- Moving coil meter
- Moving iron meter
- Moving rectifier meter
Oscilloscope
Aka O’scope. Graphically displays electrical signals and how they vary with time. Used when designing, manufacturing, and repairing electrical equipment. Comes in either 1 or 2 or 4 channels. Each channel is used to connect a probe using BNC connector. Vertical section controls the voltage or current scale for each channel. Horizontal section controls the time scale.
Passive Voltage Probe
Measures voltage difference between ground clip and hook clip. Hook is covered with a retractable cover. The BNC connector includes a capacitor compensation trimmer.
Each channel inside the oscilloscope includes a
Most probes include a switch between 1x and 10x, which refers to how much
the amplitude of the signal is attenuated (reduced). The 10x setting has an
additional
Calibration
- Set 10x on the probe
- Connect it to the oscilloscope
- Turn on the appropriate channel only
- Set Coupling setting to DC
To see all parts of the signal - Set Probe setting to 10x
- Set Type setting to Edge
- Set Source setting to the appropriate channel
- Attach ground clip to the ground tab
- Hook the probe around the signal tab
- Adjust the vertical and horizontal scale knobs
To clearly see the waveform. Turning the knobs clockwise decreases the time or voltage scale (like zooming in). - Adjust the smaller knob of the appropriate channel
To center the waveform vertically - Adjust the Trigger level betwen min and max of the waveform.
- Adjust the capacitor compensation trimmer
Until the square wave has straight edges
While visualizing 2 voltages of a circuit, the ground clip is connected to the reference point. Only 1 ground clip of the multiple probes must be connected to prevent short circuits inside the oscilloscope.
Current Probe
Operation is similar to current transformer. Steps down the current, converts to a voltage signal, and renders it in the oscilloscope. Has a sensor head, which can be opened by the opening lever. Powered through power supply cable.
Calibration
- Connect the BNC connector to a channel
- Connect power supply cable into a power supply port in the rear side of the oscilloscope
- Connect the sensor head to the circuit
- Turn on the channel
- Adjust knobs in vertical and horizontal sections